Biological Evaluation Process / Biocompatibility Testing

Process of Biological Evaluation / Biocompatibility Testing

10/15/20151 min read

Dr. Monica Cameo | BioAplicada Director | Buenos Aires, Argentina

BIOLOGICAL EVALUATION PROCESS ACCORDING TO ISO 10993

1. Material characterization (ISO 10993-18)

2. Biological evaluation tests
a) General concepts (ISO 10993-1)
b) Description of tests (ISO 10993-3, 4, 6, 9, 10, 11, 13, 14, 15)

3. Interpretation of results and biological safety evaluation

Material characterization (ISO 10993-18)

  • This is the first step in the biological evaluation process and is crucial.

  • The scope of chemical characterization required will depend on:
    a) The relevance of the information available regarding the chemical safety of the biological product and
    b) The time and nature of patient contact.

  • At a minimum, the chemical components of the medical device and any additive residues used in manufacturing must be taken into account.

Biological evaluation tests

  • Evaluate all available information and compare it with the available test data.

  • Identify the data gaps and/or additional tests required to complete the information (Annex A, Table 1).

PRIORITIES (ISO 10993-1)

1st priority: protection of human health.

2nd priority: ensure animal welfare and minimize the number and exposure of laboratory animals (3Rs).

IN VIVO TESTING

Additional in vivo tests should not be performed when:

1. Reliable results from previous tests are available.

2. Preclinical and clinical studies exist, along with a history of “safe use.”

When evaluating the relevance of previous data for the Biological Evaluation of a Medical Device (MD), ISO 10993-18 Annex C provides basic principles for assessing toxicological equivalence.

List of biocompatibility tests

(ISO 10993-1, 6.2) 6.2.2.2 Cytotoxicity (in vitro)

6.2.2.3 Delayed hypersensitivity

6.2.2.4 Irritation (including intracutaneous reactivity)

6.2.2.5 Systemic toxicity (acute)

6.2.2.6 Subacute and subchronic toxicity

6.2.2.7 Genotoxicity (in vitro)

6.2.2.8 Implantation

6.2.2.9 Hemocompatibility (in vitro)

6.2.2.10 Chronic toxicity

6.2.2.11 Carcinogenicity

6.2.2.12 Reproductive and developmental toxicity (in vitro)

6.2.2.13 Biodegradation

6.2.2.14 Toxicokinetics

6.2.2.15 Immunological toxicity

TEST SELECTION

1) Contact or exposure (nature, degree, duration, frequency, and conditions)

2) Physical and chemical nature of the final product

3) Toxicity of the chemical components in the final formulation

4) Relationship between the surface area of the medical device and body mass

5) Sensitivity and specificity of the test being considered

Additional Biological Evaluation Tests

a) Tests must be performed on the final product, sterilized.

b) Sensitive, precise, and accurate methods must be used, and testing must be conducted in accordance with GLP when applicable.