Biocompatibility of Elastomeric Stoppers
Biocompatibility of elastomeric stoppers. Elastomeric stoppers are a component of the packaging that may come into contact with the packaged drug, which is intended to enter the patient's body. Therefore, it is essential to evaluate their biocompatibility.
4/28/20171 min read


Elastomeric stoppers are commonly used to seal pharmaceutical containers. They may be of natural or synthetic origin and are generally made from a complex mixture of various ingredients, including a base polymer, vulcanizing agents, accelerators, pigments, etc.
As with any Medical Device (MD), their properties depend not only on their components but also on the manufacturing process and all the compounds used during it, including molding, mixing, cleaning, disinfection, packaging, sterilization, etc.
Elastomeric stoppers are a component of the packaging that may come into contact with the packaged drug, which is intended to enter the patient's body. Therefore, it is essential to evaluate their biocompatibility.
To evaluate biocompatibility (the ability of a medical device to demonstrate acceptable biological reactivity during its direct or indirect contact with the body), the USP indicates that in vitro tests should be performed first. Unlike what happens with other MDs, in some cases where the requirements of the in vitro tests are met, additional tests will not be necessary. Otherwise, in vivo tests will be required.
The Agar Diffusion Test or Direct Contact Test, in combination with the Elution Test, is the in vitro testing protocol recommended by the USP for plastics and polymers.
In the Agar Diffusion Test, the cells exposed to the material being tested are protected from potential mechanical damage by a layer of agar, which must also allow the diffusion of leachable substances from the polymeric material and the study of their effects at the cellular level.
In cases where we need to test, for example, due to the large size of the MD, an extract of the material will be applied to an inert filter paper, which will then be supported on the agar layer.
— Source USP<381>, <1031>, <87>, <88>, FA 7ª Edición